Compound Imaging Technology and Echogenic Needle Design: Effects on Needle Visibility and Tissue Imaging
  1. Thomas Wiesmann, MD,
  2. Andreas Bornträger, MD,
  3. Martin Zoremba, MD,
  4. Martin Neff, MD,
  5. Hinnerk Wulf, MD and
  6. Thorsten Steinfeldt, MD
  1. Department of Anesthesiology and Intensive Care Medicine, University Hospital Marburg, Marburg, Germany.
  1. Address correspondence to: Thomas Wiesmann, MD, Department of Anesthesiology and Intensive Care Medicine, University Hospital Marburg, Baldingerstrasse 1, 35033 Marburg, Germany (e-mail: wiesmann{at}med.uni-marburg.de).

Abstract

Introduction Needle visualization in ultrasound-guided regional anesthesia can be improved by using needles of echogenic design with higher rate of reflection of ultrasound waves. Imaging solutions such as compound imaging might further improve imaging of both needle and tissue; these effects have not yet been studied. We hypothesized that compound imaging would significantly improve needle visibility, regardless of the insertion angle or needle type used. The effects of compound imaging on needle artifacts and tissue imaging were also investigated.

Methods A total of 200 video clips of in-plane needle insertions were obtained in embalmed cadavers with a conventional needle and an echogenic needle at 5 different insertion angles, with both conventional B-mode ultrasound imaging and compound imaging technology. Visibility of the needle shaft and needle tip as well as the needle artifact rate were assessed by a blinded investigator on a 4-point ordinal scale. The effects on tissue image quality and speckle artifacts were also assessed. Stepwise linear regression was performed to differentiate effects on needle visibility scores.

Results Imaging of the needle shaft and tip was significantly enhanced when compound imaging technology was used (P < 0.0001). Use of echogenically designed needles or shallow needle insertion angles improved visibility of both shaft and tip (both P < 0.0001). With compound imaging, there are fewer needle artifacts, and tissue imaging quality and speckle artifact rate are significantly improved.

Conclusions Compound imaging technology enhances needle imaging with both echogenic and conventional needles. Tissue imaging and speckle artifacts are also optimized. Echogenic needle design results in better needle visibility scores in both B-mode and compound imaging.

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • This study was performed with institutional grants.

    Dr Wulf has received speaker fees and/or research funding by Sonosite, B. Braun, and Teleflex Medical (Arrow). Dr Steinfeldt has received speaker fees and/or research funding by Sonosite, B. Braun, and Teleflex Medical (Arrow).

    Supplemental digital content is available for this article. Direct URL citations appear in the printed text and are provided in the HTML and PDF versions of this article on the journal’s Web site (www.rapm.org).