Epidural infections, bacteriostatic drug effects and technical strategies for prevention
  1. Joerg Bruenke1,
  2. Thomas Riemann2,
  3. Paul Kessler3 and
  4. Norman Kachel2
  1. 1 Dr. Brünke MTC e.K, Nuremberg, Germany
  2. 2 B. Braun Melsungen AG, Melsungen, Hessen, Germany
  3. 3 Department of Anesthesiology, University Hospital Frankfurt, Frankfurt, Germany
  1. Correspondence to Dr Joerg Bruenke, Dr. Brünke MTC e.K., Nuremberg 90480, Germany; jb{at}bruenke-mtc.de

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Introduction

Epidural infections are a rare but serious complication which can result in an irreversible neurological damage if not treated. Catheterization of the epidural space is routinely used in anesthesia and analgesia, which connects the epidural space with the outside environment, through which bacteria can gain access to the epidural space.1–3 Here, the use of epidural filter serves as a physical barrier to prevent ingress of bacteria through the infusion line. In this study, we investigated the survivability of seven typical pathogens in bupivacaine and ropivacaine when contaminated during manipulation (eg, local anesthesia (LA) bolus application/infusion) and the ability of an epidural flat filter to prevent microbial transmission by a contaminated infusion solution until physical occlusion of the filter.

Methods

The survivability of seven typical device-associated pathogens was tested by artificial microbial contamination (1.25×105 colony-forming units (CFU)/mL) of different concentrations of two commonly used epidural anesthetics, bupivacaine and ropivacaine. The number of vital micro-organisms was checked daily over a period of 4 days by plating out aliquots and CFU counting.

The maintenance of the microbial barrier function of a 0.2 µm epidural catheter filter (Perifix from B. Braun) was investigated by the presence of microbial contamination in the flow-through solution after challenging the filter with an artificially contaminated infusion solution with test bacteria Staphylococcus aureus or Pseudomonas aeruginosa by a volumetric infusion pump at a speed of 10 mL/hour until automated pressure shutdown was observed due to physical/mechanical occlusion of the filter by bacteria.

Results

Microbial survivability in anesthetics bupivacaine and ropivacaine

The number of vital organisms in the anesthetics bupivacaine and ropivacaine was reduced over time. However, some pathogens (S. aureus, P. aeruginosa and Candida albicans) were able to survive to a certain degree over the test period of 4 days (figure 1).

Figure 1

Viability of microorganisms in the anesthetic solutions bupivacaine and ropivacaine over a period of 4 days; the survivability of micro-organisms in the anesthetics, bupivacaine and ropivacaine, respectively, was tested using seven different microbial strains, often found in device-associated infections. The spectrum of species included four Gram-positive strains (Staphylococcus epidermidis DSM 18857, S. aureus MRSA DSM 21979, Enterococcus faecalis DSM 2570/ATCC 29212 and Streptococcus pneumoniae DSM 11867), two Gram-negative bacterial strains (Pseudomonas aeruginosa DSM 939/ATCC 15442 and Escherichia coli DSM 1576/ATCC 8739) and one yeast (Candida albicans DSM 5817/ATCC 10259). The vitality of all test organisms in the anesthetic solutions bupivacaine (0.5%, 0.25% and 1.75% solution) and ropivacaine (0.5% and 0.25% solution) was determined every day over a period of 4 days by serial dilution plating of a 1 mL aliquot of the inoculated anesthetic solution. The data were from two individual experiments and normalized to the number of vital bacteria at test start (d0).

Microbial tightness of epidural filter

For physical occlusion of a 0.2 µm epidural flat filter by micro-organisms, a relatively high number of bacteria (≥5.5×108 CFUs) were required for both tested micro-organisms, S. epidermidis and P. aeruginosa, respectively. The filter maintained its microbial barrier function even after physical occlusion as no bacteria were detectable in the flow-through fraction (figure 2). The level of vital bacteria remained comparable throughout the test.

Figure 2

Microbial tightness testing: microbial evaluation of flow-through samples. The epidural flat filter was challenged with an infusion solution artificially contaminated with ~2.0×106 CFU/mL until the automated infusion system shut down to overpressure due to physical occlusion caused by the test bacteria in the infusion solution. At different time points (test start, after 250 mL and after physical occlusion), aliquots from the solution that passed the epidural flat filter were plated out on plate count agar plates, and the number of CFUs was determined after incubation at 37°C for a day. To prove the vitality of the test organism throughout the test procedure, the number of bacteria was also determined in the infusion solution at test start and after the filter was physically occluded. Physical occlusion of the 0.2 µm epidural flat filter was achieved at a relatively high number of bacteria, 5.5×108 CFU for Pseudomonas aeruginosa DSM 939 and 9.5×108 CFU for Staphylococcus epidermidis DSM 18857, respectively. CFU, colony-forming unit.

Discussion

This in vitro study demonstrated that different typical device-associated pathogens can survive over a period of several days directly in the anesthetic solutions bupivacaine and ropivacaine. Furthermore, it was shown that an 0.2 µm epidural flat filter can efficiently prevent microbial transmission even when challenged with high numbers of bacteria until physical occlusion and pressure shutdown was achieved.

It was shown that ropivacaine exhibited poor antibacterial activities,4–6 whereas bupivacaine showed concentration-dependent antibacterial activity.4 While other studies investigated on antibacterial activity in the presence of growth medium, our study focused on the ability to survive directly in the anesthetic solution simulating microbial contamination during drug admixture.

Maintaining filter integrity and microbial tightness even when physical occlusion by microorganism is achieved has not been reported so far. Interestingly, the number of micro-organisms leading to physical occlusion of the epidural filter was observed with a very high concentration of bacteria (>108 CFU).

Noteworthy, the contamination level chosen in this study was higher than that in a normal clinical environment in order to determine the required bioburden to obtain physical occlusion of the epidural filter. The results of our investigation demonstrate that microbial barrier function was even sustained when physical occlusion of the peridural flat filter by >108 bacteria was attained. Thus, the duration time of epidural flat filters is not limited by time but mainly by the flow rate, which is influenced by the number of germs, particles and air which can block the membrane mechanically. Thus, the recommendation is to keep the system closed as long as technically possible (closed system). However, this study investigated selected aspects of the epidural filter in an in vitro setup and, thus, the behavior in an in vivo clinical environment could be different and has to be further investigated in clinical settings.

Ethics statements

Patient consent for publication