Ultrasound Characteristics of Needles for Regional Anesthesia
  1. Tim Maecken, M.D.,
  2. Michael Zenz, M.D., Ph.D. and
  3. Thomas Grau, M.D., Ph.D.
  1. Clinic for Anaesthesiology, Intensive Care, Palliative Care & Pain Medicine, BG University Clinics Bergmannsheil, Bochum, Germany.
  1. Reprint requests: Tim Maecken, M.D., Clinic for Anaesthesiology, Intensive Care, Palliative Care & Pain Medicine, Bürkle-de-la-Camp Platz 1, D-44789 Bochum, Germany. E-mail: maecken@anaesthesia.de/tim.maecken{at}ruhr-uni-bochum.de

Abstract

Background and Objectives: Needle visibility is a crucial requirement for successful and safe ultrasound-guided peripheral nerve blocks. We performed a prospective study on the ultrasonic imaging quality of various commercially available needles. We tested the hypothesis that different nerve block needles would have different ultrasonic appearances. Furthermore, we examined the influence of needle angle with regard to the ultrasound plane, 2 types of media surrounding the needle, and the impact of 3 different ultrasound machines.

Methods: Twelve needles were prospectively tested for 3 ultrasound planes (longitudinal, axial tip, and axial shaft) at 2 different angles (0 degrees and 45 degrees). Quality of needle visibility and ultrasound scans were described by using 6 criteria (visibility score range 0-10): (1) visibility of needle; (2) visibility of surrounding media; (3) consistency of needle surface; (4) formation of artifacts; (5) degree of shadowing; and (6) detection and distinction of the needle from the surrounding media. Additionally, every ultrasound scan was performed in 2 media (water bath and animal model) with 3 ultrasound devices and evaluated by 2 investigators. Evaluation of the ultrasound scans was blinded with regard to needle but not to the ultrasound machine and media.

Results: In the animal model, visibility was good at 0 degrees (visibility score greater than 6) but was decreased for all needles at a 45 degree angle (criterion 6). In this setting, 2 needles were difficult to identify (score less than 3; criterion 6) and only 3 of 12 reached a score of 7 or more (criterion 6). Depiction quality for all 3 planes was significantly lower in the animal model when compared with the water bath (P < .001) and at an angle of 45 degrees when compared with 0 degrees (P < .001). There was no significant impact of the ultrasound machine on image quality.

Conclusions: In a tissue-equivalent model we found significant differences among different types of needles at a 45 degree angle. In clinical use, angles between 30 degrees and 60 degrees are required. Because visibility of the needle is a keystone of ultrasound-guided peripheral nerve blocks, our results suggest the need to optimize the echogenicity of needles used for ultrasound-guided nerve blocks.

  • Ultrasound
  • Regional anesthesia
  • Needle visibility

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • The study group has been involved in lecturing, education, and training that has been supported by the following companies: Sonosite, GE Healthcare, Esaote, B. Braun Melsungen and Pajunk. B. Braun funded the expense of all tested needles as well as hardware for video recordings. The study group received financial and technical support by the ISURA group (International Society of Ultrasound and Regional Anesthesia) and institutional funds of the BG University Clinics, Bergmannsheil, Bochum, Germany.