Article Text

Download PDFPDF
Determination of Epidural Catheter Placement Using Nerve Stimulation in Obstetric Patients
  1. Ban C.H. Tsui, M.S.C., M.D.,
  2. Sunil Gupta, M.D., F.R.C.P.C. and
  3. Brendan Finucane, M.B.B.Ch., F.R.C.P.C.
  1. From the Department of Anaesthesia, University of Alberta and affiliated hospitals (UAH and Grey Nuns Hospital), Edmonton, Alberta, Canada.
  1. Reprint requests: Ban C.H. Tsui, Department of Anaesthesia, University of Alberta Hospitals, 3B2.32 Walter Mackenzie Health Sciences Centre, 8440–12 Street, Edmonton, Alberta, Canada T6G 2B7.

Abstract

Background and Objectives Peripheral nerve and spinal cord stimulation techniques have been used for many years. However, electrical stimulation methods rarely have been used to confirm epidural catheter placement. This study examines the practicality of this technique to confirm epidural catheter placement in obstetric patients.

Methods Thirty-nine obstetric patients in labor were studied. An electrocardiography (ECG) adapter (Arrow-Johans) was attached to the proximal end of the epidural catheter (19-gauge Arrow Flextip plus). Then, the lowest milliamperage (1–10 mA) necessary to produce a motor response (truncal or limb movement) was applied using this setup. A positive motor response indicated that the catheter was in the epidural space. This test was performed initially after catheter insertion and repeatedly after incremental local anesthetic doses.

Results The sensitivity and specificity of the new test was 100% and 100%, respectively, with 38 true positive tests and 1 true negative test. The threshold current required to produce a positive test increased after incremental doses of local anesthetic. A case of intravascular epidural catheter migration was detected using this new test. In this case, the patient's increased threshold milliamperage returned to the baseline value within 1 hour, despite having received local anesthetic. Intravascular epidural catheter migration was subsequently confirmed by a positive epinephrine test dose. Thus, this test appears to be a potentially useful method to detect intravascular catheter placement.

Conclusion This study demonstrates that this test may have a role in improving the success rate of epidural anesthesia. Reg Anesth Pain Med 1999: 24: 17–23.

  • epidural
  • catheter placement
  • test
  • nerve stimulation.

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • This study was presented in part at the 23rd Annual Meeting American Society of Regional Anesthesia, Seattle, Washington, May 1998.