Article Text

Download PDFPDF
Analysis of Human Epidural Pressures
  1. Sebastian P. Thomas, M.D.*,
  2. John I. Gerson, M.D.** and
  3. Gerald Strong, M.D.
  1. From the Department of Anesthesiology, SUNY Health Science Center, Syracuse, New York.
  2. *Associate Professor of Anesthesiology and Director, Pain Treatment Service.
  3. **Associate Professor of Anesthesiology, SUNY Health Science Center.
  4. Attending Anesthesiologist, St. Vincent's Medical Center, Jacksonville, Florida.
  1. Address correspondence to P. Sebastian Thomas, M.D., Department of Anesthesiology, SUNY Health Science Center, 750 E. Adams St., Syracuse, NY 13210.

Abstract

Background and Objectives. During performance of epidural injection, entry into the epidural space has traditionally been determined by identification of negative pressure to the advancing needle by indirect means such as hanging drop; loss of resistance to air, saline, or water; or use of a MacIntosh balloon. Confusion in the literature regarding entry pressures versus postentry (baseline) pressures and zero reference pressure was noted.

Methods. Baseline epidural pressure changes were examined using a closed system zeroed to the dorsal spine during and after injection of local anesthetics in 39 patients referred to the Pain Treatment Center, located at SUNY Health Science Center, Syracuse, New York, for epidural injections. The course of epidural pressure changes after injection and the temporal relation of epidural pressure waves to the arterial and venous wave forms were examined.

Results. In contrast to previous investigations, subatmospheric pressure was found in the epidural space in only one patient. Baseline pressure for all patients was 7.7 ± 3.9 mmHg. There were significant differences in baseline pressure in patients who had undergone back surgery compared with patients who had not undergone such surgery: 11.8 ± 3.4 as opposed to 7.0 ± 3.5 mmHg, respectively ( p < 0.005). Three minutes after a 2-ml injection of local anesthetic into the epidural space, the pressure returned to baseline. This contrasted to the effects of a 6-ml injection, which resulted in the pressure remaining above baseline after 3 minutes. The epidural pressure wave forms more closely paralleled the radial artery wave form than the central venous pressure.

Conclusion. Lumbar epidural pressure is greater than atmospheric pressure when referenced to zero at the dorsal spine level.

  • Epidural block
  • baseline epidural pressure
  • postlaminectomy.

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • Performed at SUNY Health Science Center, Syracuse, New York.

    The authors acknowledge the assistance of Dr. Peter Byles in planning the project, Ms. Fran Kriese for technical support. Dr. Richard Oates for advice on data analysis, and Ms. Diane Coe and Ms. Anne Marie Fabian for editorial assistance.